Preparation of Artificial CSF
Artificial cerebrospinal fluid (CSF) more closely mirrors the composition of endogenous CSF, and is therefore more physiologically compatible. Following is a method for preparing artificial CSF.
These references can offer key information on agent stability, solubility,
pharmacokinetics, and techniques for optimum delivery. Citations in
the bibliography are indexed by the criteria shown here. The information in the ALZET
pump bibliography is made available as a complimentary service. Dozens
of popular reference lists can be found in the ALZET bibliography.
1. Preparation of Solution A
Dissolve in 500 ml pyrogen-free, sterile water:
Compound |
(g) |
NaCl |
8.66 |
KCl |
0.224 |
CaCl2 · 2H2O |
0.206 |
MgCl2 · 6H2O |
0.163 |
2. Preparation of Solution B
Dissolve in 500 ml pyrogen-free, sterile water:
Compound |
(g) |
Na2HPO4 · 7H2O |
0.214 |
NaH2PO4 · H2O |
0.027 |
3. Preparation of Artificial CSF
Combine solutions A and B in a 1:1 ratio.
4. Composition of Cerebrospinal Fluid in Various Species 1
Ion/Compound |
Human |
Dog |
Cat |
Rabbit |
Na |
187.5 |
153.5 |
160 |
149* |
K |
2.6 |
3.1 |
4.4 |
2.9* |
Ca |
1.1 |
1.4 |
1.4 |
1.24* |
Mg |
1.1 |
1.0* |
0.7* |
0.9* |
P |
0.8 |
0.4* |
-- |
-- |
Cl |
119 |
122 |
147 |
99 |
HCO3 |
23.3* |
25.8* |
24.6* |
22* |
(Concentration in mM)
*Note : Only one reference listed a concentration level. Therefore, the listed value was used.
1 Ion concentrations (except for the levels indicated with an asterisk) are an average of the data listed in Davson, H. Physiology of the Cerebrospinal Fluid, J. & A. Churchill, Ltd., London, 1967 and Biology Data Book , Volume III, 2nd ed., Fed. Am. Soc. Exper. Biol., Washington D.C., 1974
5. Comparison of Real and Artificial CSR Electrolyte Concentrations
Ion/Compound |
Cerebrospinal Fluid* |
Artificial CSF |
Na |
154 |
150 |
K |
3.0 |
3.0 |
Ca |
1.4 |
1.4 |
Mg |
0.9 |
0.8 |
P |
0.4 |
1.0 |
Cl |
136 |
155 |
HCO3 |
24.1 |
-- |
(Concentration in mM)
*Note : Cerebrospinal fluid concentrations are an average of the values listed for the various animal species shown in #4 above.
1 Ion concentrations (except for the levels indicated with an asterisk) are an average of the data listed in Davson, H. Physiology of the Cerebrospinal Fluid, J. & A. Churchill, Ltd., London, 1967 and Biology Data Book , Volume III, 2nd ed., Fed. Am. Soc. Exper. Biol., Washington D.C., 1974
6. Explanation
for the Absence of HCO3
Bicarbonate was not added to the formula for artificial CSF for two reasons:
- Bicarbonate can cause shifts in the pH of the solution as it converts to CO2.
- As HCO3converts to CO2 , bubbles can develop inside the pump. The presence of a gas in the pump can then affect the pumping rate in unpredictable ways.
Therefore, unless there is an urgent need for using bicarbonate in a study, we strongly recommend that it not be added to the formula for artificial CSF.
top of page